1. Measurement principle of flue dust transmittance detector
The flue dust transmittance detector is also called the flue dust monitor. The measurement principle of the flue dust transmittance detector is that the detection light traverses the flue. The single optical path measurement system consists of a transmitter and a receiver. A high-brightness LED light source is installed in the receiver. It emits a beam that passes through the diffuser and then passes through the flue. It is projected onto the focusing lens in the receiver and is focused on the measurement detector; the other part is directly focused on the reference detector. The light transmittance or light opacity value can be obtained by calculating the ratio of the signal intensities obtained on the two detectors, from which the light transmittance (or light opacity) and dust concentration values can be obtained.
2. Main features of flue dust transmittance detector
direct continuous measurement
Small size, light weight and compact structure
Direct display of light transmittance or opacity
With lens contamination monitoring and compensation device
Automatic calibration and manual calibration, convenient and flexible
Measuring path 0.5-5 meters
With air purge channel
Long life LED signal light source
No moving parts – low maintenance
Simple installation and debugging
3. Technical parameters of flue dust transmittance detector
working principle | Across the flue, single optical path transmission |
light source | High brightness LED |
Response time | 3 seconds to reach 98% of final value |
Measuring range | Light transmittance 0-99.99% |
Resolution | 0.01% |
Detectable gas | Smoke, fumes or related dust gases |
Optical path length | 0.3-2 meters standard type (other lengths can be customized) |
Average time | 1-59 seconds |
Calibration and Calibration | Manual calibration and calibration (with calibrator) |
show | LCD liquid crystal display |
Analog output | 4-20mA current loop |
digital output | RS485/232 |
Relay output | Two switching outputs (alarm value setting) |
Operating temperature | -20-50°C |
Working humidity | 0-99% (no condensation) |
Supply voltage | 220VAC; optional 12, 24VDC |
Power consumption | Transceiver system maximum 18W |
Vibration | 2Hz-60Hz, maximum amplitude 2mm |
Shell material | die cast aluminum |
Protection level | IP67 |