1. COD Analyzer principle
COD Analyzer uses potassium dichromate method and photometric method, suitable for municipal sewage, industrial water and surface water monitoring.Add a certain amount of potassium dichromate solution to the sample, use silver sulfate as a catalyst in a strong acidic medium, use a unique dilution injection technology to reduce the chloride ion concentration, and use a joint masking agent to mask high-concentration chloride ions, and then pass through high temperature and high pressure After digestion, the COD concentration value was determined using spectrophotometry.
2. Product features of COD Analyzer
The new photoelectric quantitative technology is not affected by interference such as large chroma, suspended solids, and bubbles. It has high reliability, accuracy, and strong anti-interference ability.
The detection accuracy is high, the lower limit of measurement is low, and the long-term drift of the instrument is small.
It has one-click self-test function and self-diagnosis function.
Using a single-channel highly integrated valve group, maintenance and cleaning are simple and fast.
All-weather networking function allows you to monitor the operating status of the instrument anytime and anywhere.
It has alarm function, quality control function and reverse control function.
Equipped with automatic cleaning function and automatic calibration function.
Equipped with whole point measurement, interval measurement and external control measurement.
It will work automatically after power failure and recovery.
Automatic prompts and automatic reset functions for faults and lack of reagents (samples).
With fault recording function.
Can save more than 3 years of historical data.
Meet the requirements of separate discharge of cleaning liquid and waste liquid
3. COD Analyzer technical parameters
Measurement method: According to the standard HJ828-2017 "Water Quality-Chemical Oxygen Consumption Determination-Dichromate Method"
Measuring range: 0-200 mg/L; 0-500 mg/L; 0-1000 mg/L; range can be customized
Indication error: ≤±3% Detection limit: 10mg/L
Repeatability: ≤±3% Waste liquid volume: 6mL/time
Memory effect: ≤±3% Actual water sample comparison: ±8%
Constant temperature time: 900 seconds Voltage stability: ≤±5%
Measurement period: 39 minutes Calibration period: any specified time
Maintenance cycle: ≥720h/time Maintenance workload: <2 hours/month
Reagent consumption: 3 months/500ml Display output: Equipped with 10.1-inch color LCD touch screen
Color temperature: The recommended temperature is 165°C, which can be set according to the actual water sample conditions
Data export: Measured values can be imported into a U disk and saved through the USB port.
Signal output: RS485/RS232/USB interface/standard configuration: 2 channels of 4-20mA output/standard configuration of two channels of switching input and output
Environmental requirements: Temperature-controlled indoors, recommended temperature (5-28°C), humidity ≦90% (no condensation)
Power supply and power: (220V±22)V/AC, (50±0.5Hz), 5A, 150W
Instrument size: upper cabinet 600*450*300mm; upper cabinet 700*450*300mm